Bedienungsanleitung de FA 200 - 2 Drucktaupunkt - Testo AG

EinbettenHerunterladen
FA 200 - 2
Drucktaupunkt-Messgerät
Pressure dew point measuring instrument
Bedienungsanleitung
de
Instruction manual
en
Inhalt
Seite
Vorwort................................................................................................................................2
Sicherheitshinweise ............................................................................................................3
Beschreibung ......................................................................................................................3
Geräteabbildung ..................................................................................................................4
Geräteabmessungen ..........................................................................................................4
Installation ..........................................................................................................................5
Direkt im Druckluftnetz ................................................................................................5
Indirekt im Druckluftnetz ..............................................................................................5
Einstellungen ......................................................................................................................6
Schnellübersicht ..........................................................................................................6
Messen................................................................................................................................7
Elektrischer Anschluss ................................................................................................7
Gehäuse öffnen ..........................................................................................................7
1. Messgröße einstellen ..............................................................................................7
2. Kalibrieren / Abgleich ..............................................................................................7
3. Alarm einstellen ......................................................................................................8
4. Skalieren..................................................................................................................9
5. Werkseinstellung......................................................................................................9
Taupunktdiagramm für Druckluft ........................................................................................10
Technische Daten ..............................................................................................................11
Bestelldaten ......................................................................................................................13
Vorwort
Liebe Testo Kundin,
lieber Testo-Kunde,
Ihre Entscheidung für ein Messgerät von Testo war richtig. Jedes Jahr kaufen tausende
Kunden
unsere hochwertigen Produkte. Dafür sprechen mindestens 7 gute Gründe:
1)
2)
3)
4)
5)
6)
Bei uns stimmt das Preis-Leistungs-Verhältnis. Zuverlässige Qualität zum fairen Preis.
Deutlich verlängerte Garantiezeiten von bis zu 3 Jahren - je nach Gerät!
Mit der fachlichen Erfahrung von über 40 Jahren lösen wir Ihre Messaufgabe optimal.
Unser hoher Qualitätsanspruch ist bestätigt durch das Zertifikat nach ISO9001.
Selbstverständlich tragen unsere Geräte das von der EU geforderte CE-Zeichen.
Kalibrier-Zertifikate für alle relevanten Messgrößen. Seminare, Beratung und
Kalibrierung vor Ort.
7) Auch nach dem Kauf lassen wir Sie nicht im Regen stehen.
Unser Service garantiert Ihnen schnelle Hilfe.
Dieses Produkt erfüllt laut Konformitätsbescheinigung die Richtlinien
gemäß 89/336/EWG
2
Sicherheitshinweise
Vor Inbetriebnahme lesen!
Achtung: Netzspannung 230 V auf Relais-Kontakt.
Absicherung max. 1 A bauseits bei angeschlossener Abschirmung.
Druckbereich >50 bar nicht überschreiten.
Messbereiche das Messwertaufnehmers beachten!
Bei Überhitzung werden die Fühler zerstört.
Zulässige Lager- und Transporttemperatur sowie die
zulässige Betriebstemperatur beachten
(z. B. Messgerät vor direkter Sonneneinstrahlung schützen).
Bei Öffnen des Geräts, unsachgemäßer Behandlung
oder Gewaltanwendung erlöschen die Gewährleistungsansprüche!
Einstell- und Kalibrierarbeiten nur durch qualifiziertes
Personal aus der Mess- und Regeltechnik durchführen lassen.
Bei transienten Störgrößen (Elektrostatische Entladung sowie Burstimpulse) kann es zu einem Reset des Messsystems kommem.
Beschreibung
Prozesssicherheit in Druckluftsystemen
Das Einhalten eines bestimmten Feuchtegehalts (Drucktaupunkt) ist für jedes Druckluftsystem die Grundvoraussetzung für einen dauerhaft störungsfreien Anlagenbetrieb. Die
hohen Qualitätsstandards in der modernen Industrieproduktion erfordern einen kontinuierliche Feuchteüberwachung mit Alarmmeldung.
Das FA 200-2
Der Drucktaupunkt tpd ist die Temperatur, bei der die Druckluft gesättigt ist. Das FA 200-2
liefert ein Ausgangssignal 4...20 mA entsprechend -60...+30 °C tpd für die Weiterverarbeitung. Das LED-Display zeigt den herrschenden Drucktaupunkt. Über die Tastatur kann
zusätzlich ein Alarm/Grenzwert programmiert und die relative Feuchte eingestellt werden.
Bei Grenzwertüberschreitung schaltet ein Relaiskontakt. Zusätzlich blinkt das Display.
Vorteil:
Überwachung und Alamierung rund um die Uhr.
3
Geräteabbildung
LED-Anzeige
Sinterkappe
G 1/2”-Gewinde
Abschirmung
Potentialfreier Kontakt
230 V / 1 A
Stromausgang
4...20 mA
(+) blau
(-) schwarz
FunktionsTasten
Spannungsversorgung
24 V DC
(+) weiss
(-) braun
Common:
NO:
NC:
PE:
(Belegung siehe Seite 8)
Geräteabmessungen (in mm)
51
12
76
52
141
192
44
50
57
schwarz
blau
braun
gelb/grün
4
Installation
Direkt im Druckluftnetz
Druckluftleitung
FA 200-2
Fühler mit dem G 1/2”-Gewinde druckdicht mittig oder
oben in die zu messende Druckluftleitung einschrauben.
Darauf achten, dass dicht am Druckluftstrom gemessen
wird. Bei Sackleitungen und nicht strömender Druckluft
ergeben sich sehr lange Reaktionszeiten für den Feuchtemesswert. Es empfiehlt sich nach Trocknen der Druckluft
und allen Bypassleitungen oder auch bei kritischen
Druckluft-Verbrauchern die Installation durchzuführen.
Indirekt im Druckluftnetz
Druckluftleitung
Messkammer
Kapillarleitung
FA 200-2
Normanschluss
Fühler mit dem G 1/2”-Gewinde in die Messkammer einschrauben. Messkammer verbinden mit der Druckluftleitung über einen Kugelhahn und eventuell eine diffusionsdichte Anschlussleitung (max. 5 m). Bei öl- und
schmutzhaltiger Druckluft einen Vorfilter 40 µm vor der
Messkammer installieren. Über die Kapillarleitung der
Messkammer strömt kontinuierlich etwas Druckluft ab
(bei 7 bar ca. 1 l/min expandiert). Die Reaktionszeiten
für den Feuchtemesswert sind kürzer als bei der direkten
Montage.
5
Einstellungen
Schnellübersicht
Tasten
Text
↑
↓
C
S
Drücken/Sek. Blinken/Hz
1. Messgröße einstellen
Öffnen
1 = °C tpd (Werkseinstellung)
2 = °F tpd
3 = %rF
4 = °Ct
5 = °Ft
Escape
Speichern
x
x
2
x
x
2
2
x
2
2. Kalibrieren/Abgleich
Kalibrieren
im Modus
- °C tpd, °F
- % rF
Escape
Speichern
x
2 sec. on
0,5 sec. off
tpd
x
x
x
2
2
x
2
2
x
2
2
2
3. Alarm einstellen
Einstellung
Speichern
Alarm
Auslesen Anzahl Alarm
Escape
Speichern (zurück)
x
x
x
x
2 sec. on
0,5 sec. off
3
4. Skalieren
4mA
20 mA
Escape
Speichern
x
x
x
x
x
x
2
2
2
2
x
2
x
5. Werkseinstellung
Werkseinstellung
x
x
6
x
2 sec. on
0,5 sec. off
Messen
Elektrischer Anschluss
Falsches Anschließen führt zum Defekt des Sensors.
Der potentialfreie Kontakt ist mit 230 V / 1 A abzusichern.
Bei Inbetriebnahme eine Spannung von 24 V DC an-schließen. Im Display blinkt ca. 4 sec. der mittlere Balken. Es erscheint der zuletzt gewählte Modus in der
Anzeige.
Gehäuse öffnen
Vorsichtig Deckel abschrauben.
1. Messgröße einstellen
↑
↓
Wählbare Messgößen sind:
Feuchte
1 °C tpd
2 °F tpd
3 %rF
Umgebungstemperatur
4 °Ct
5 °Ft
↓”
Einstellen aktivieren:
Taste ”↓
↑” oder Taste”↓
↓”
Messgröße (1 - 5) auswählen:Taste ”↑
Auswahl speichern:
Taste ”S” 2 sec. drücken.
↑”
Auswahl abbrechen:
gleichzeitig die Tasten ”↑
↓” 2 sec. drücken.
und ”↓
C
S
2. Kalibrieren / Abgleich
(bezieht sich auf die Messgröße im Display)
↑
↓
Das FA 200-2 kann vor Ort kalibriert bzw. abgeglichen
werden. Mögliche Einstellungen sind relative Feuchte oder
Taupunkt.
2.1 Anzeige °C tpd oder °F tpd (Drucktaupunkt)
(Abgleich mit Referenzgerät)
Einstellen aktivieren:
C
S
Messwert einstellen:
Auswahl speichern:
Auswahl abbrechen:
7
gleichzeitig Taste ”S” und
Taste ”C” 2 sec. drücken.
Display blinkt.
↑” oder Taste”↓
↓”
Taste ”↑
Taste ”S” 2 sec. drücken.
↑”
gleichzeitig dieTasten ”↑
↓” 2 sec. drücken.
und ”↓
Messen
Kalibrieren / Abgleich
2.2 Anzeige %rF (Abgleich mit Kontrollset 11,3 %rF)
Hinweis
- Vor jeder Kalibrierung die Sinterkappe vorsichtig abschrauben.
- Sensor nicht berühren.
- Erschütterungen vermeiden.
- Verschmutzte Sensoren und Sinterkappe mit reinem
Alkohol oder destiliertem Wasser reinigen.
- Sinterkappe von innen nach aussen mit Druckluft ausblasen.
- Kontrolle durchführen nur mit aufgeschraubten Kontroll-/
Abgleichtopf 11,3 %rF (Best.-Nr. 0699.3397).
- Angleichzeit 30 Minuten.
- Temperaturschwankungen und Sonneneinstrahlung vermeiden.
↓
↑
Ist die Abweichung zum Kontrollset >2%, Abgleich durchführen.
Einstellen aktivieren:
Wert speichern:
Auswahl abbrechen:
3. Alarm einstellen (Überschreiten)
S
C
gleichzeitig Taste ”S” und
Taste ”C” 2 sec. drücken.
Display blinkt mit 2 sec. ON,
0,5 sec OFF
Taste ”S” 2 sec. drücken.
↑”
gleichzeitig dieTasten ”↑
↓” 2 sec. drücken.
und ”↓
(bezieht sich auf die Messgröße im Display)
Es besteht die Möglichkeit den oberen Grenzwert einzu-
blau
schwarz stellen.
braun
Bei Überschreiten des eingestellten Grenzwerts schaltet
das Alarmrelais. Das Display blinkt 3x pro sec.
Bei Unterschreiten blinkt das Display nicht.
NC
NO
C
PE
Schutzleiter
gelb/grün
↑
↓
Einstellen aktivieren:
Grenzwertert einstellen:
Einstellen abbrechen:
Grenzwert speichern:
Auslesen Anzahl Alarm:
Zurück ins Messmenü:
C
S
8
↑”
Taste ”↑
↑” oder Taste”↓
↓”
Taste ”↑
↑”
gleichzeitig dieTasten ”↑
↓” 2 sec. drücken,
und ”↓
alter Wert bleibt gespeichert.
Taste ”S” 2 sec. drücken.
Taste ”C” 2 sec. drücken.
Anzahl der Überschreitungen wird angezeigt.
Taste ”S” 2 sec. drücken.
Speicherinhalt wird auf 0
gesetzt.
Messen
4. Skalieren
Standardeinstellung ab Werk (je nach Displayauswahl)
-60...+30 °C tpd
-76...+86 °F tpd
± 0...100 %rF
-20...+50 °C
- 4...+122 °F
= 4...20 mA
= 4...20 mA
= 4...20 mA
= 4...20 mA
= 4...20 mA
Eingestellte Alarm-Grenzwerte ab Werk:
+30 °C tpd
+86 °F tpd
100 %rF
+50 °C
+122 °F
Vor dem Skalieren Messgröße einstellen (siehe Seite 7)
↑
↓
4 mA
Aktivieren:
Wert einstellen:
Speichern:
Abbrechen:
20 mA
Aktivieren:
C
↑
S
Wert einstellen:
Speichern:
Abbrechen:
↓
gleichzeitig Taste ”S” und
↓” 2 sec. drücken.
Taste ”↓
Display blinkt.
↑” oder Taste”↓
↓”
Taste ”↑
Taste ”S” 2 sec. drücken.
↑”
gleichzeitig dieTasten ”↑
↓” 2 sec. drücken.
und ”↓
gleichzeitig Taste ”S” und
↑” 2 sec. drücken.
Taste ”↑
Display blinkt.
↑” oder Taste”↓
↓”
Taste ”↑
Taste ”S” 2 sec. drücken.
↑”
gleichzeitig dieTasten ”↑
↓” 2 sec. drücken.
und ”↓
5. Werkseinstellung
Aktivieren:
gleichzeitig Tasten ”S”, ”C”,
↑” und ”↓
↓” 2 sec. drüc”↑
ken.
Das Gerät ist auf Werkseinstellungen zurückgesetzt.
C
S
9
Taupunktdiagramm für Druckluft
100
g
) *
m3
(
Die Druckreduzierung an
der Entnahmestelle führt
physikalisch bedingt zu
einer Reduzierung des
Drucktaupunkt-Wertes.
40 35 30
10
7
5
4
65
60
55
50
45
15
3
70
20
20
40
25
30
70 °65 60 55 50 45
50
Gramm Wasserdampf je m3 feuchtigkeitsgesättigter Druckluft
70
Dieses Diagramm gibt Aufschuss über die
Drucktaupunkt-Veränderung bei Druckverlust. Als Beispiel ist ein Druckverlust von
8 bar auf 6 bar Betriebsüberdruck dargestellt. Der Drucktaupunkt wandert in
diesem Fall von 10 °C auf 5 °C ab.
20
-5
15
0
-1
10
-1
2¡
5
5
0,3
25
0
0,4
30
2
0,5
5
0,7
10
1,0
35
0¡
0
-2
0,2
Drucktaupunkt °C
40
2
-5
5
-2
0
-3
-10
0,1
Luftdruck (bar)
1
2
3
4
5
7
10
14
20
Luftdruck (bar)
0
1
2
3
4
6
9
13
19
* bezogen auf 0 bar, 20 °C
10
Technische Daten
FA 200-2
Messbereich:
-60...30 °C tpd
0...100 %rF umschaltbar
Druckbereich:
-1,0...bis 50 bar
Genauigkeit:
±2 °C tpd
bei -5 °C tpd typisch
Auflösung:
1 °C
Stromvesorgung:
24 VDC, 3.0 VA verwechslungssicher +10 % und -15 % UN
Stromausgang:
4...20 mA (galvanisch getrennt)
=-60...+30 °Ctd bzw. 0...100 %rF
max. Bürde:
< 500 Ω
Anzeige:
3stellig LED-Display
Schutzart:
IP 65
Anschluss
(mechanisch):
G 1/2” Einschraubgewinde
(Edelstahl)
Anschluss
(elektrisch):
Flanschstecker/
Kupplungsdose M 12 x 1
24 V DC (+) weiss
(-) braun
4...20 mA (+) blau
(-) schwarz
Feuchtigkeit:
<75% im Jahresmittel
Relaiskontakt:
1 Wechsler potentialfrei
(230V AC, 1 A)
Grenzwert:
frei einstellbar über
Tastatur
Einsatztemperatur:
0...50 °C
Lager-/Transporttemperatur:
-20...70 °C
Abmessungen (in mm):
(B x H x T) 50 x 67 x 197
11
Technische Daten
FA 200-2
Kalibrierung:
über 4 interne Tasten in
1°C tpd-Schritte oder bei
11,3%rF
Anzeige:
7 Segment LED
Alarmfunktion:
Grenzwerteinstellung über 4 interne Tasten, LED-Display blinkt
bei Grenzwertüberschreitung
Potentialfreier
Alarmausgang:
Wechsler 230 V / 1A belastbar
elektrischer Anschluss
Flanschstecker/
Kupplungsdose M 12 x 1
4 x 0,35 mm2 Kabel
incl. Schutzleiter
Common:
NO:
NC:
PE:
schwarz
blau
braun
gelb/grün
Skalierung
Stromausgang:
Skalierung über 4 interne
Tasten für 4...20 mA
Anzeige verschiedener
Werte:
Bedienung über 4 interne Tasten
1
°C tpd
2
°F tpd
3
%rF
4
°C t
5
°F t
Auslesen Alarm:
die Anzahl aufgetretener
Alarme kann ausgelesen werden
Garantie:
2 Jahre
12
Bestelldaten
Bestelldaten
Artikel-Nr.
FA 200-2, Drucktaupunktmessgerät .................................. 0556.2002
Messkammer mit Standardanschluss..................................0554.3303
Anschlussleitung Länge 5 m, 4polig mit Abschirmung ......0699.3391
Anschlussleitung Länge 10 m, 4polig mit Abschirmung ....0699.3392
Anschlussleitung Alarmkontakt Länge 5 m,
4polig incl. Schutzleiter ........................................0699.3393
Anschlussleitung Alarmkontakt Länge 10 m,
4polig incl. Schutzleiter ........................................0699.3394
Netzteil (Tischgerät) 90...264VAC/24VDC (3A) ....................0554.1748
ISO-Kalibrierprotokoll bei -40 und -10°Ctpd (6bar) ..............0520.0136
Für den mobilen Einsatz
testo 645 inkl. TopSafe, Batterie und
Bedienungsanleitung............................................0563.6450
Drucktaupunktfühler ..........................................................0636.9840
Anschlussleitung, Verb. Fühler ↔ Messgerät ....................0430.0143
Transport-Koffer ................................................................0516.3250
®
testo -Protokolldrucker......................................................0554.0545
13
FA 200 - 2
Pressure dew point measuring instrument
Bedienungsanleitung
de
Instruction manual
en
Contents
Page
Introduction ........................................................................................................................2
Notes on safety ..................................................................................................................3
Description ..........................................................................................................................3
Diagram of instrument ........................................................................................................4
Dimensions of instrument ....................................................................................................4
Installation ..........................................................................................................................5
Directly in the compressed air system ........................................................................5
Indirectly in the compressed air system ......................................................................5
Settings ..............................................................................................................................6
Quick overview ............................................................................................................6
Measuring............................................................................................................................7
Electrical connection....................................................................................................7
Opening the housing ..................................................................................................7
1. Setting the parameter..............................................................................................7
2. Calibrating / Adjusting ............................................................................................7
3. Setting the alarm ....................................................................................................8
4. Scaling ....................................................................................................................9
5. Factory setting ........................................................................................................9
Dew point diagram for compressed air..............................................................................10
Technical data ..................................................................................................................11
Ordering data ....................................................................................................................13
Introduction
You have made the right decision by choosing a measuring instrument from Testo.
Thousands of customers buy our high standard products every year. There are at least 7
good reasons for doing so:
1) Cost-performance ratio. Reliable quality at a fair price.
2) Extended warranty times of up to 3 years - depending on instrument.
3) We have the ideal solutions for your measuring tasks based on our expert experience
gained over 40 years.
4) Our high quality standard is confirmed by the ISO 9001 certificate.
5) Of course, our instruments carry the CE symbol required by the EU.
6) Calibration certificates for all relevant parameters. Seminars, advice and calibration on
location.
7) Our after-sales service. Ask for more details.
The conformity certificate confirms that this product fulfills the guidelines in accordance
with 89/336/EWG.
2
Notes on safety
Please read prior to operation
Warning: Mains voltage of 230 V at relay point.
Fuse protection of max. 1 A by the customer if shield is attached.
Do not exceed pressure range of >50 bar.
Observe measuring ranges of sensor!
The probes are damaged if overheated.
Observe max. storage and transport temperature and
max. operating temperature
(e.g. protect measuring instrument from direct sunlight).
Warranty claims no longer apply if the instrument is opened, in the case
of inexpert handling or use of force.
Adjustments or calibrations should be carried out by qualified
measurement and control engineering staff.
Transient disturbances (electrostatic discharge as well as bursts)
can lead to a reset of the measuring system.
Description
Process stability in compressed air systems
Adherence to a specific moisture level (pressure dew point) is required of every compressed
air system to ensure long-lasting uninterrupted operation. The high quality standards in
modern industrial production require non-stop monitoring of moisture with alarm.
The FA 200-2
Pressure dew point tpd is the temperature at which the compressed air is saturated.
FA 200-2 supplies an output signal of 4 to 20 mA corresponding to -60 to +30 °C tpd for further processing. The current pressure dew point is shown in the LED display. The keypad is
also used to program the alarm/limit value and to set relative humidity.
A relay point is switched when the limit value is exceeded. The display also flashes.
Advantage:
Round-the-clock monitoring and alarm.
3
Diagram of instrument
LED display
Sintered cap
G 1/2” thread
Cover
Floating contact
230 V / 1 A
Power output
4 to 20 mA
(+) blue
Common:
NO:
NC:
PE:
(-) black
Function
buttons
Power supply
24 V DC
(+) white
(-) brown
(See page 8 for
assignment)
Dimensions of instrument (in mm)
51
12
76
52
141
192
44
50
57
black
blue
brown
yellow/green
4
Installation
Directly in the compressed air system
Compressed air pipe
FA 200-2
Compressed air
pipe
Measuring
chamber
Capillary
pipe
FA 200-2
Standard
connection
Screw in probe with G 1/2 thread pressure-tight in the
centre or in the compressed air pipe where the
measurement is to take place.
Ensure that the measurement is carried out close to the
compressed air flow. U-bend pipes or non-flowing
compressed air result in very slow reaction times for the
moisture reading. Installation is recommended following
drying of the compressed air and all bypass pipes or for
critical compressed air users.
Indirectly in the compressed air system
Screw in probe with the G 1/2” thread in the measuring
chamber. Connect measuring chamber with the
compressed air pipe using a ball valve and possibly a
diffusion-tight connection pipe (max. 5m). In the case of
compressed air containing oil and dirt particles, a 40 µm
pre-filter should be installed in front of the measuring
chamber. Compressed air flows continuously
(at 7 bar, approx. 1 l/min. expanded) in the capillary pipe of
the measuring chamber. The reaction times for the humidity
reading are shorter than when directly mounted.
5
Settings
Quick overview
Buttons
Text
↑
↓
C
S
Press/sec.
Flash/Hz
1. Setting the parameter
Open
1 = °C tpd (Factory setting)
2 = °F tpd
3 = %RH
4 = °Ct
5 = °Ft
Escape
Save
x
x
2
x
x
2
2
x
2
2. Calibrating/Adjusting
Calibration
in mode
- °C tpd, °F
- % RH
Escape
Save
x
2 sec. on
0.5 sec. off
tpd
x
x
x
2
2
x
2
2
x
2
2
2
3. Setting the alarm
Setting
Save
Alarm
Read out Number Alarm
Escape
Save (restore)
x
x
x
x
2 sec. on
0.5 sec. off
3
4. Scaling
4mA
20 mA
Escape
Save
x
x
x
x
x
x
2
2
2
2
x
2
x
5. Factory setting
Factory setting
x
x
6
x
2 sec. on
0.5 sec. off
Measuring
Electrical connection
Incorrect connection cause defects in the sensor.
The floating contact is protected with 230 V / 1 A.
Connect power of 24 V DC for initial operation. The middle
bar flashes for approx. 4 sec. in the display. The mode last
selected is shown.
Opening the housing
Unscrew the cover carefully.
1. Setting the parameter
↑
↓
The following parameters can be selected:
Moisture
1 °C tpd
2 °F tpd
3 %RH
Ambient temperature
4 °Ct
5 °Ft
Activate setting:
Select parameter (1 - 5):
Save selection:
Cancel selection:
C
S
↓” button
“↓
↑” button or “↓
↓” button
“↑
Press “S” for 2 sec.
↑” and “↓
↓” buttons
Press the “↑
simultaneously for 2 sec.
2. Calibrating / Adjusting
(refers to the parameter in the display)
↑
↓
FA 200-2 can be calibrated or adjusted on location.
Relative humidity or dew point are possible settings.
2.1 Display: °C tpd or °F tpd (pressure dew point)
(Adjust using reference instrument)
Activate setting:
C
S
Set reading:
Save selection:
Cancel selection:
7
Press “S” and “C” buttons
simultaneously for 2 sec.
Display flashes.
↑” button or “↓
↓” button
“↑
Press “S” for 2 sec.
↑” and “↓
↓” buttons
Press the “↑
simultaneously for 2 sec.
Measuring
Calibrating / Adjusting
2.2 Display: %RH (adjustment with 11.3 %RH control set)
Note
- Carefully remove the sintered cap prior to every
calibration
- Do not touch sensor.
- Avoid vibrations.
- Clean dirty sensors and sintered cap with pure alcohol or
destilled water.
- Flush sintered cap from the inside to the outside with
compressed air.
- Check should only be carried out using screwed-on
11.3%RH control/adjustment container
(Part no. 0699.3397).
- Adaptation time: 30 minutes.
- Avoid fluctuations in temperature and sunlight.
If the deviation from the control set >2% adjustment must
be carried out.
↓
↑
Activate setting:
Press “S” and “C” buttons
simultaneously for 2 sec.
Display flashes with 2 sec.
ON, 0.5 sec OFF.
Press “S” for 2 sec.
↑” and “↓
↓” buttons
Press “↑
simultaneously for 2 sec.
Save value:
Cancel selection:
brown
3. Setting the alarm (exceeding)
S
C
(refers to the parameter in the display)
blue
PE
Protective
conductor
yellow/green
↑
C
NC
NO
black
↓
It is possible to set the top limit value.
The alarm relay switches when the set limit value is
exceeded. The display flashes 3x/sec.
The display does not flash if the value undercuts the limit.
Activate setting:
Set limit value:
Cancel setting:
Save limit value:
Read out Quantity Alarm:
is displayed.
Return to measurement
menu
C
S
8
↑” button
“↑
↑” button or “↓
↓” button
“↑
↑” and ”↓
↓” buttons
Press the “↑
simultaneously for 2 sec.,
old value remains saved.
Press “S” for 2 sec.
Press “C” for 2 sec.
Number of times exceeded
Press “S” for 2 sec.
Memory contents are set
at 0.
Measuring
4. Scaling
Standard factory setting (depending on display selection)
-60 to +30 °C tpd
-76 to +86 °F tpd
± 0 to 100 %RH
-20 to + 50 °C
- 4 to +122 °F
= 4 to 20 mA
= 4 to 20 mA
= 4 to 20 mA
= 4 to 20 mA
= 4 to 20 mA
Alarm limit values set in factory:
+30 °C tpd
+86 °F tpd
100 %RH
+50 °C
+122 °F
Set parameter before scaling (see page 7)
↑
↓
Activate 4 mA:
Set value:
Save:
Cancel:
simultaneously for 2 sec.
Activate 20 mA:
C
↑
S
Set value:
Save:
Cancel:
↓
S
↑” buttons
Press “S” and “↑
simultaneously for 2 sec.
Display flashes.
↑” button or “↓
↓” button
“↑
Press “S” button for 2 sec.
↑” and “↓
↓” buttons
Press “↑
simultaneously for 2 sec.
5. Factory setting
↑” and “↓
↓”
Press “S”, “C”, “↑
buttons simultaneously for
2 sec.
The instrument is reset to
the factory settings.
Activate:
C
↓” buttons
Press “S” and “↓
simultaneously for 2 sec.
Display flashes.
↑” button or “↓
↓” button
“↑
Press “S” for 2 sec.
↑” and “↓
↓” buttons
Press the “↑
9
Dew point diagram for compressed air
100
g
) *
m3
(
A reduction in pressure at
the sampling point may
result in a reduction of
the pressure dew point
value.
40 35 30
10
7
5
4
65
60
55
50
45
15
3
70
20
20
40
25
30
70 °65 60 55 50 45
50
Gramme of water vapour per m3 of moisture saturated compressed air
70
The diagram provides information on the
change in pressure dew point when there is
a drop in pressure. Example: a drop in
pressure from 8 bar to 6 bar working
positive pressure is shown. In this case the
pressure dew point drops from 10 °C to 5
°C.
20
-5
15
0
-1
10
-1
2¡
5
5
0,3
25
0
0,4
30
2
0,5
5
0,7
10
1,0
35
0¡
0
-2
0,2
Pressure dew point °C
40
2
-5
5
-2
0
-3
-10
0,1
Air pressure (bar) 1
2
3
4
5
7
10
14
20
Air pressure (bar) 0
1
2
3
4
6
9
13
19
* with reference to 0 bar, 20 °C
10
Technical data
FA 200-2
Measuring range:
-60 to 30 °C tpd
0 to 100 %RH, switchable
Pressure range:
-1.0 to 50 bar
Accuracy:
±2 °C tpd
at -5 °C tpd typically
Resolution:
1 °C
Power supply:
24 VDC, 3.0 VA, cannot be
confused, +10 % and -15 % UN
Power output:
4 to 20 mA (electrically isolated)
=-60 to +30 °Ctd or 0 to 100 %RH
Max. load
< 500 Ω
Display:
3 digit LED display
Protection class:
IP 65
Connection
(mechanical):
G 1/2” screw-in thread
(stainless steel)
Connection
(electrical):
Flange plug/
coupling socket M 12 x 1
24 V DC (+) white
(-) brown
4 to 20 mA (+) blue
(-) black
Moisture:
<75% on average
Relay point:
1 change-over contact, floating
(230V AC, 1 A)
Limit value:
Freely selectable via keypad
Application temperature: 0 to 50 °C
Storage/transport
temperature:
-20 to 70 °C
Dimensions (in mm):
(w x h x d) 50 x 67 x 197
11
Technical data
FA 200-2
Calibration:
Via 4 internal buttons in
1°C tpd steps or at
11.3%RH
Display:
7 segment LED
Alarm function:
Limit value setting via 4 internal
buttons, LED display flashes
when limit value is exceeded
Floating
alarm output:
Change-over contact, can take
230 V / 1A
Electrical connection
flange plug/
coupling socket M 12 x 1
4 x 0.35 mm2 cable
incl. protective conductor
Common:
NO:
NC:
PE:
black
blue
brown
yellow/green
Scaling
power output:
Scaling via 4 internal
buttons for 4 to 20 mA
Display of different
values:
Operation via 4 internal buttons
1
°C tpd
2
°F tpd
3
%RH
4
°C t
5
°F t
Read off alarm:
can be read off.
The number of activated alarms
Warranty:
2 years
12
Ordering data
Ordering data
Item no.
FA 200-2, pressure dew point measuring instrument .......... 0556.2002
Measuring chamber with standard connection ..................0554.3303
Connection cable 5 m long, 4 pin with cover ....................0699.3391
Connection cable 10 m long, 4 pin with cover ..................0699.3392
Connection cable alarm contact, 5 m long,
4 pin with protective conductor ..........................0699.3393
Connection cable alarm contact, 10 m long,
4 pin with protective conductor ..........................0699.3394
Power supply unit 90...264VAC/24VDC (3A) ......................0554.1748
ISO-Calibration-Certificate at -40 and -10°Ctpd ..................0520.0136
For mobile use
testo 645 with TopSafe, battery and
instruction manual................................................0563.6450
Pressure dew point probe ................................................0636.9840
Cable, connects probe ↔ measuring instrument ................0430.0143
Transport case....................................................................0516.3250
®
testo printer ......................................................................0554.0545
13
testo AG
Postfach 11 40, 79849 Lenzkirch
Testo-Straße 1, 79853 Lenzkirch
Telefon: (07653) 681-0
Fax: (07653) 681-100
E-Mail: info@testo.de
Internet: http://www.testo.com
0973.5002/01/T/wh/19.03.2004